Select Seller
You‘ve never treated teeth this gently before: DentaPort ZX with Optimum Torque Reverse function (OTR) ensures optimal torque of the file – both in rotary and alternating mode. Mircocracks and file breakage are a thing of the past. This is endodontics of the highest class – especially for you.
Gentle and safe root canal preparation with OTR function – the new Optimum Torque Reverse function
The new OTR function unites the advantages of the rotating and cyclical-reciprocating technologies: as long as the adjusted torque value is not exceeded, the motor rotates permanently and, hence, advances at an optimal rate. As soon as the pressure on the file permanently exceeds the set value during 180° rotation, the Torque Reverse will be activated and the rotation of the file will be reversed by 90° to release the instrument. Then the file will rotate in the cutting direction again.
Endometric measurement, canal preparation and polymerization in one system
You don’t just want reliable measurements, you also want root canal preparation with visual monitoring and comfortable finishing? Simply adjust the DentaPort ZX OTR to suit your requirements. With its modular design, the combination system for convenient and safe endodontics turns the device into exactly what you want whenever you need it:
Reliable apex localization with DentaPort Root ZX
The apex locator, the base module of the DentaPort ZX is characterized by the following features:
Automatic calibration feature – even when canal conditions change!
With its light and simultaneously sturdy housing, the successor to the legendary Root ZX is arguably the best-selling apex locator worldwide – and its excellent references are further proof of its high quality.
Comfortable preparation with DentaPort TriAuto OTR
When added to DentaPort Root ZX the DentaPort TriAuto OTR becomes a comfortable endodontic motor with measuring function. The respective position of the file is shown on the display while you are preparing the root canal with the compact handpiece.
Fast finishing with Curing light
Simply replace the motor handpiece with the curing handpiece in order to switch over to adhesive treatment. This high-performance module shows you all parameters clearly on a large display.
Integrated file electrode in the handpiece or more working length
Denta Port’s new contra angle has a markedly smaller head diameter and is fitted with a built-in file electrode for more working length.
You should replace the file electrode after about 100 operating hours (rotation time). DentaPort ZX will emit an acoustic signal as a reminder.
Automatic safety functions
You can rely on the integrated safety system: the DentaPort ZX Set OTR always knows what to do. The file starts to rotate as soon as it enters the root canal – and stops again as soon as it is removed. And the automatic safety functions are activated as soon as file rotation gets close to the individually set torque limits or the reference point.
Automatic start-stop-function
Motor begins automatically when the file is inserted in the root canal and stops when it is withdrawn. This ensures a safe and effective preparation.
Active protection against file breakage
Optimum Torque Reverse
The file rotates in the cutting direction without strain. If the preset torque is exceeded, it alternates 180° forwards and 90° back.
Very small torque values for the torque limitation can be set by means of this function.
Auto Torque Slow-down
Auto Torque Slow-down The rotational speed is reduced as soon as the file approaches the preset torque limit. This prevents torsional fatigue.
Auto Torque Reverse
The file stops automatically upon reaching a specific torque and changes its rotational direction.
Auto Apical Slow-down
The rotational speed is reduced as soon as the file approaches a previously set reference point to the apex. This increases the safety in the critical apical area.
Auto Apical Stop
The file stops automatically upon reaching a stipulated reference point.
Auto Apical Reverse
The file stops automatically and rotates in the opposite direction as soon as a stipulated reference point has been reached.
This prevents excessive instrumentation and apex perforation.
Only 1-3 files are needed for the preparation procedure
Benefit from the simplified work processes: You will need only one and to three files at the most to safely prepare the root canal and can use all customary file systems, except the Reciproc system.
You save both material and time. After all, the safe removal of debris, improvement of canal access and retention of the original canal contour are all achieved with a significantly lower expenditure of time.
Existing devices always can be upgraded with the OTR function
DentaPort TriAuto OTR is available as separate module. Existing devices always can be upgraded with the OTR function.
Configuration:
Technical Data :
Product name
DentaPort TriAuto OTR Modul
Current supply:
Metal-Hydride DC 9.6 V, 20 VA
Motor:
0 – 800 rpm.
Dimensions (mm):
W 105 x D 127 x H 115
Weight:
approx. 430 g
Weight handpiece:
onyl 70 g